Ferromagnetic Glass-Coated Micro-and Nanowires for Medical Application

SA Baranov1,2*

Author Affiliations

1Institute of Applied Physics, Academiei str. 5, Chisinau, MD 2028 Republic of Moldova, Russia

2Shevchenko Pridnestrov’e State University, str. 25 Oktyabrya 128, Tiraspol, Republic of Moldova, Russiam

Received: December 23, 2020 | Published: January 06, 2021

Corresponding author: SA Baranov, Institute of Applied Physics, Academiei str. 5, Chisinau, MD 2028, Shevchenko Pridnestrov’e State University, str. 25 Oktyabrya 128, Tiraspol, Republic of Moldova, Russia

DOI: 10.26717/BJSTR.2021.32.005329

Abstract

The properties of shielding made of a glass-coated amorphous magnetic micro- and nanowires have been studied by radio noise suppression experiments. This effect can be used also for contactless diagnostics of deformations in distant objects (including organs of the body) reinforced by cast magnetic microwires with the stress effect. The absorption parameter values observed are of (5-20) dB at (0,1-10) GHz. The losses profile is ascribed to the presence of natural ferromagnetic resonance in microwires. The natural ferromagnetic resonance (NFMR) reveals large residual stresses appearing in the microwire core in the course of casting. These stresses, together with the magnetostriction, determine the magnetoelastic anisotropy. Beside the residual internal stresses, the NFMR frequency is influenced by external stresses applied to the microwire or to the composite containing the latter (the so-called stress effect). The dependence of the NFMR frequency on the deformation of the microwires is proposed to be used in the distant diagnostics of deformations of medical objects (including organs of the body).

Keywords: Glass-coated Amorphous Magnetic Micro-and Nanowires; Magnetostriction; Magnetoelastic Anisotropy; Natural Ferromagnetic Resonance

Introduction

The increasingly miniaturized devices at high frequencies lead to noise in medical apparats. This fact has strongly promoted the development for noise suppression purposes [1-3]. Application magnetic glass-coated cast amorphous micro and nanowires (GCAMNWs) is considered also in articles [1-4]. The properties of the magnetic glass-coated cast amorphous micro and nanowires studies in many publications by various research groups [1-16]. As this research has evolved, the need to control the magnetic anisotropy in components came into attention since the field absorption is related to natural ferromagnetic resonance (NFR) [5- 10]. That glass-coated amorphous magnetic micro-and nanowires (GCAMNWs) possess very interesting high frequency properties due to the strong magnetoelastic anisotropy which is induced in the fabrication process [5]. By defining strictly, the geometric characteristics and the alloy composition the medical norms of electromagnetic field can be accurately controlled. Furthermore, the high production rate and high permeability feature of GCAMNWs without any rigid substrates are advantageous for use as noise absorbers of electromagnetic field. In this report, amorphous glasscoated microwire arrays in a sheet configuration are proposed for use as absorbers. High frequency characteristics of thin amorphous microwires will be presented and evaluated as radio noise absorbers and medical protection radio absorption shielding.

During the manufacturing, the GCAMNWs using the Ulitovsky- Taylor method the soften glass capillary is filled with a melted metal alloy with a further quenching [5]. The residual stresses, arising in metal core of GCAMNWs by during manufacturing, determine magnetic behavior of such GCAMNWs [5]. The metal is frozen into very thin GCAMNWs. Thus, a strong quenching stress (for example, s ~109 Pa in ultrathin GCAMNWs) is introduced in the GCAMNWs [4-12]. The phenomenon of natural ferromagnetic resonance (NFMR) in CGCMMWs [5-14] is extremely interesting from the viewpoint of using it for non-contact diagnostics of distant objects (or organs of the body). The diagnostics become possible due to the stress effect on the NFMR, that is, the shift of the resonance frequency as a result of a deformation of the object (or organs of the body). Such a frequency shift can be measured by irradiating the object with microwaves emitted by radar at frequencies near the NFMR and detecting the reflected signal, thus revealing a deviation of the resonance frequency from the original value. The glass coating of the cast GCAMNWs induces strong mechanical stresses in the kernel [4-12]. In cylindrical coordinates, the residual tension is characterized by the axial, z σ , radial, r σ , and tangential, ϕ σ , components. The value of these stresses depends on the ratio of the radius, Rm, of the metallic kernel to the total microwire radius, Rc:

Using the cylinder-shell model, we then obtain a formula for stresses in the metallic kernel of the cast GCAMNWs:

is the maximum stress in the metallic kernel; ε is the difference between the thermal expansion of the metallic kernel and that of the glass shell with the expansion coefficients α1 and α2: ε = (α1 – α2)(T* – T); E1 is the Young modulus of the metallic kernel, T* is the solidification temperature of the composite in the metal /glass contact region (T* ~ (800…1200) K), T is the room temperature; k is the ratio between Young’s moduli of the glass and the metal ; ν is the Poisson ratio. Let us consider the case where all the Poisson ratios are ν = 1/3 in order to obtain

With additional longitudinal strain, which occurs when the microwire is embedded in a solid matrix that itself deforms under external influence, the following term is added to the expression for the residual axial tension:

Where Po is the force applied to the composite; Sm= π (Rm)2 is the microwire cross-sectional area. The theory of NFMR is presented in Refs. [5-10]. The NFMR frequency can be written as

Design of Composites for Radio-Absorption Shielding

The Designs of Composites from GCAMNWs have Following Configurations

However, for very thin GCAMNWs with negative, positive and zero magnetostriction the domain structure is far more complicated due to a complex stress distribution within the small diameter. The simplest domain structure reported so far for these GCAMNWs is obtained in [10,12]. As for the microwire array configuration for noise suppression experiments, 223 microwires (10 μm total diameter and length of 5 mm where the metal core diameter is 2 μm,) are parallel aligned (Figures 1a & 1b) in an area of 5 x 4 mm2 as can be observed in (Figure 1c). They are supported by a thin polymer sheet (  100 μm thickness) [1-3].

Figure 1a: Composite shielding for radio absorption protection with GCAMNWs where were maked in grating form. We can consider two types of orientation of a magnetic field: E and H.

Figure 1b: Micrograph of an array of Co based microwires.

Figure 1c: Composite shielding for radio absorption protection with GCAMNWs where were made in a stochastic mixture of microwires in the polymeric matrix.

Noise Absorption Experiments

In order to investigate the noise absorption performances a microstrip line (with 50 Ω characteristic impedance) has been designed [1-3]. Similar equipment can be used to study the stress effect on the NFMR. The signal line has a width of (6-10) μm and thickness of t = 35 μm and is set on a 300 thick Teflon base with permittivity r ε = 3.6. The insertion losses are estimated below 0.08 at 1GHz. The performance of the line is measured using an Agilent E8363B network analyzer from 0.1 to 20 GHz. The microwire array was placed on the top of the signal line. Samples of a single and twofold array of microwires were fabricated. The measured absorption coefficient S21 on the microstrip line is showed in [1]. The S21 parameter at 1 GHz decreases from the background of -0.08 dB to -0.154 dB and -0.188 dB for the single layer and twofold layer (labeled as A to B). It is worth mentioning that larger decreases are observed at around 20GHz. This is probably due to the metallic property of the wires. The ratio of the power loss by the microwire layers to the input power calculated as Ploss/Pin = 1-(|S11|2+|S21|2) is shown in (Figure 2) [1]. A loss peak nearby the resonance frequency is observed around for both samples. It is noted that the twofold sample fails to be effective correspondingly. The reason is that the second layer is not close enough to interact with the near field from the microstrip line.

Figure 2: Power loss profile of Co based GCAMNWs shielding.

From the experimental results described above the following matters should be considered for future improvements. Although the amorphous wires exhibit superb frequency profiles and various advantages such as a high production rate and geometrical flexibility, the metal cores in the wires are coated with very thick glass layer resulting in a poor packing density of high permeability region in the wire array samples. If the packing density of 0.12 for the present samples were, for example 0.8, the power loss would be about 0.3 for a single layer. This is possible by removing the glass coating. Furthermore, by designing a wire array with proper conductivity and permeability, a high-performance absorber. These issues will be treated intensively as this research work progresses in near future.

Conclusion

The high permeability features for very thin amorphous microwires for medical apparats and HF radio absorption shielding were investigated. The electromagnetic noise absorption of sheet type samples with dimensions (5 x 4) mm2 made of 223 parallel aligned microwires was measured. The two-fold layers exhibit of -2.5 dB, with insertion losses below 0.2 at 1 GHz which corresponds to the resonance of magnetization at (1-10)GHz [1-3]. On the other hand, we have presented simple analytical expressions for the residual and external stresses in the metallic kernel of the microwire, which clearly show their dependence on the ratio of the external radius of the microwire to the radius of the metal kernel and on the ratio of Young’s modules of glass and metal. The NFMR phenomenon observed in glass-coated magnetic microwires opens up the possibility of developing new materials with a wide range of properties [6-9]. An important feature of cast microwires with an amorphous magnetic core is the dependence of the NFMR frequency from the deformation [7,8]. Therefore, this effect can be used for contactless diagnostics of deformations in distant objects (including organs of the body) reinforced by cast magnetic microwires with the stress effect. These objects are periodically scanned with floating-frequency radar to determine the deviation of the initial NFMR frequency due to potentially dangerous deformations of the monitored object.

Another principle of detecting mechanical strain is examined in [15,16]. This principle is based on the giant magnetoimpedance (GMI) effect. The GMI effect [15,16] demands external magnetization of the sample which is not required in the NFMR method [6-9].

References

Review Article

Ferromagnetic Glass-Coated Micro-and Nanowires for Medical Application

SA Baranov1,2*

Author Affiliations

1Institute of Applied Physics, Academiei str. 5, Chisinau, MD 2028 Republic of Moldova, Russia

2Shevchenko Pridnestrov’e State University, str. 25 Oktyabrya 128, Tiraspol, Republic of Moldova, Russia

Received: December 23, 2020 | Published: January 06, 2021

Corresponding author: SA Baranov, Institute of Applied Physics, Academiei str. 5, Chisinau, MD 2028, Shevchenko Pridnestrov’e State University, str. 25 Oktyabrya 128, Tiraspol, Republic of Moldova, Russia

DOI: 10.26717/BJSTR.2021.32.005329

Abstract

The properties of shielding made of a glass-coated amorphous magnetic micro- and nanowires have been studied by radio noise suppression experiments. This effect can be used also for contactless diagnostics of deformations in distant objects (including organs of the body) reinforced by cast magnetic microwires with the stress effect. The absorption parameter values observed are of (5-20) dB at (0,1-10) GHz. The losses profile is ascribed to the presence of natural ferromagnetic resonance in microwires. The natural ferromagnetic resonance (NFMR) reveals large residual stresses appearing in the microwire core in the course of casting. These stresses, together with the magnetostriction, determine the magnetoelastic anisotropy. Beside the residual internal stresses, the NFMR frequency is influenced by external stresses applied to the microwire or to the composite containing the latter (the so-called stress effect). The dependence of the NFMR frequency on the deformation of the microwires is proposed to be used in the distant diagnostics of deformations of medical objects (including organs of the body).

Keywords: Glass-coated Amorphous Magnetic Micro-and Nanowires; Magnetostriction; Magnetoelastic Anisotropy; Natural Ferromagnetic Resonance